Chapter 4 · Moving Charges and Magnetism

4.10 The Moving Coil Galvanometer

Chapter 3 used meters to measure currents and voltages in circuits. Most analogue meters are built round one instrument, the , which uses the torque on a current loop from Section 4.9.

It has a coil of many turns, free to turn about a fixed axis, between the poles of a permanent magnet (Fig. 4.29). Inside the coil sits a cylinder of soft iron. The curved pole faces and the iron core make the field : it points along radii towards the axis, so whatever angle the coil turns to, its sides cut straight across the field. The iron also makes the field stronger.

Fig. 4.29 The moving coil galvanometer. (a) The coil, core, spring, pointer and scale. (b) Seen from above, the pole faces and core make a radial field, so the forces on the coil sides always form a couple of the same size.

With a current I in the coil, the field is always in the plane of the coil, so sin θ = 1 in Eq. (4.35) and the torque on N turns of area A is

(4.42)

A spring opposes the turning, with a restoring torque kφ for a turn through angle φ. Here k is the of the spring. The coil turns until the two torques balance:

(4.43)

so the deflection is

(4.44)

The deflection is proportional to the current, so the scale is evenly spaced (Fig. 4.30). The quantity in the bracket is fixed for a given meter.

Current through the coil (µA)

deflection 0°

0

10

20

35

50

Fig. 4.30 A galvanometer turns through 90° for a current of 50 µA. Choose a current: the deflection is proportional to it.

A galvanometer on its own cannot serve as an ammeter, for two reasons. It is very sensitive: a typical one gives a full-scale deflection for a few tens of microamperes. And its resistance, often tens of ohms, would change the very current it is meant to measure when joined in series. Both problems are solved by joining a small resistance, called a , in parallel with the coil (Fig. 4.31a). Most of the current bypasses the coil through the shunt, and the meter’s total resistance is little more than the shunt’s:

In these equations, R with the subscript G is the resistance of the galvanometer, r with the subscript s that of the shunt, and I with the subscript g the current through the coil. The result is an . If the full current I should give full-scale deflection, the potential differences across the coil and the shunt must be equal:

which gives the shunt needed:

(4.47)

Fig. 4.31 (a) An ammeter: a galvanometer with a small shunt in parallel. (b) A voltmeter: a galvanometer with a large resistance in series.

The of the galvanometer is the deflection per unit current:

(4.45)

A maker can raise it with more turns, a larger coil, a stronger magnet or a weaker spring.

To measure a potential difference, the galvanometer is joined in parallel with the part of the circuit concerned, and it should then draw almost no current. So a large resistance R is put in series with it (Fig. 4.31b), and the total resistance is close to R:

The result is a . For full-scale deflection at a potential difference V, the series resistance must be

(4.48)

The is the deflection per unit potential difference across the galvanometer itself:

(4.46)

Doubling the number of turns doubles the current sensitivity. But it also roughly doubles the length of wire, and so the resistance of the coil, which leaves the voltage sensitivity about the same.

Example 4.15

A cell of 6.0 V with negligible internal resistance is joined to a 4.0 Ω resistor. (a) A galvanometer of resistance 40.0 Ω is put in series to read the current. What does it read? (b) The galvanometer is turned into an ammeter with a shunt of 0.020 Ω. What does the ammeter read? (c) What is the current with no meter at all?

Given

Cell V = 6.0 V, resistor 4.0 Ω Galvanometer resistance 40.0 Ω Shunt 0.020 Ω

Solution

(a) The galvanometer’s own resistance adds to the circuit.

(b) The shunt and the coil in parallel make a meter of resistance

so the current is now

(c) With no meter,

The bare galvanometer cuts the current to a tenth of its true value. With the shunt, the ammeter reads within 0.5 %.

Answer

(a) 0.14 A, (b) 1.49 A, (c) 1.50 A

Example 4.16

A galvanometer of resistance 50 Ω gives full-scale deflection for a current of 2.0 mA. How would you turn it into (a) an ammeter reading up to 1.0 A and (b) a voltmeter reading up to 10 V?

Given

Galvanometer resistance 50 Ω Full-scale current 2.0 × 10⁻³ A

Solution

(a) Use Eq. (4.47) for the shunt.

(b) Use Eq. (4.48) for the series resistance.

Answer

(a) a 0.10 Ω shunt in parallel; (b) 4950 Ω in series

Check yourself

To turn a galvanometer into a voltmeter, what do you join to it?

Chapter 4 · Moving Charges and Magnetism

4.10 The Moving Coil Galvanometer

Chapter 3 used meters to measure currents and voltages in circuits. Most analogue meters are built round one instrument, the , which uses the torque on a current loop from Section 4.9.

It has a coil of many turns, free to turn about a fixed axis, between the poles of a permanent magnet (Fig. 4.29). Inside the coil sits a cylinder of soft iron. The curved pole faces and the iron core make the field : it points along radii towards the axis, so whatever angle the coil turns to, its sides cut straight across the field. The iron also makes the field stronger.

Fig. 4.29 The moving coil galvanometer. (a) The coil, core, spring, pointer and scale. (b) Seen from above, the pole faces and core make a radial field, so the forces on the coil sides always form a couple of the same size.

With a current I in the coil, the field is always in the plane of the coil, so sin θ = 1 in Eq. (4.35) and the torque on N turns of area A is

(4.42)

A spring opposes the turning, with a restoring torque kφ for a turn through angle φ. Here k is the of the spring. The coil turns until the two torques balance:

(4.43)

so the deflection is

(4.44)

The deflection is proportional to the current, so the scale is evenly spaced (Fig. 4.30). The quantity in the bracket is fixed for a given meter.

Current (µA)

deflection 0°

0

10

20

35

50

Fig. 4.30 A galvanometer turns through 90° for a current of 50 µA. Choose a current: the deflection is proportional to it.

A galvanometer on its own cannot serve as an ammeter, for two reasons. It is very sensitive: a typical one gives a full-scale deflection for a few tens of microamperes. And its resistance, often tens of ohms, would change the very current it is meant to measure when joined in series. Both problems are solved by joining a small resistance, called a , in parallel with the coil (Fig. 4.31a). Most of the current bypasses the coil through the shunt, and the meter’s total resistance is little more than the shunt’s:

In these equations, R with the subscript G is the resistance of the galvanometer, r with the subscript s that of the shunt, and I with the subscript g the current through the coil. The result is an . If the full current I should give full-scale deflection, the potential differences across the coil and the shunt must be equal:

which gives the shunt needed:

(4.47)

Fig. 4.31 (a) An ammeter: a galvanometer with a small shunt in parallel. (b) A voltmeter: a galvanometer with a large resistance in series.

The of the galvanometer is the deflection per unit current:

(4.45)

A maker can raise it with more turns, a larger coil, a stronger magnet or a weaker spring.

To measure a potential difference, the galvanometer is joined in parallel with the part of the circuit concerned, and it should then draw almost no current. So a large resistance R is put in series with it (Fig. 4.31b), and the total resistance is close to R:

The result is a . For full-scale deflection at a potential difference V, the series resistance must be

(4.48)

The is the deflection per unit potential difference across the galvanometer itself:

(4.46)

Doubling the number of turns doubles the current sensitivity. But it also roughly doubles the length of wire, and so the resistance of the coil, which leaves the voltage sensitivity about the same.

Example 4.15

A cell of 6.0 V with negligible internal resistance is joined to a 4.0 Ω resistor. (a) A galvanometer of resistance 40.0 Ω is put in series to read the current. What does it read? (b) The galvanometer is turned into an ammeter with a shunt of 0.020 Ω. What does the ammeter read? (c) What is the current with no meter at all?

Given

Cell V = 6.0 V, resistor 4.0 Ω Galvanometer resistance 40.0 Ω Shunt 0.020 Ω

Solution

(a) The galvanometer’s own resistance adds to the circuit.

(b) The shunt and the coil in parallel make a meter of resistance

so the current is now

(c) With no meter,

The bare galvanometer cuts the current to a tenth of its true value. With the shunt, the ammeter reads within 0.5 %.

Answer

(a) 0.14 A, (b) 1.49 A, (c) 1.50 A

Example 4.16

A galvanometer of resistance 50 Ω gives full-scale deflection for a current of 2.0 mA. How would you turn it into (a) an ammeter reading up to 1.0 A and (b) a voltmeter reading up to 10 V?

Given

Galvanometer resistance 50 Ω Full-scale current 2.0 × 10⁻³ A

Solution

(a) Use Eq. (4.47) for the shunt.

(b) Use Eq. (4.48) for the series resistance.

Answer

(a) a 0.10 Ω shunt in parallel; (b) 4950 Ω in series

Check yourself

To turn a galvanometer into a voltmeter, what do you join to it?